General operation method of solid phase extraction column

General operation of solid phase extraction
Reversed phase extraction column operation
Normal phase extraction column operation
Ion exchange extraction column
Extract from polar matrix
Non-polar sample
Extracted from a non-polar matrix
Polar sample
Extracting charged samples
activation
The extraction column is rinsed with 3-5 ml of methanol;
Rinse the extraction column with 3-5 ml of deionized water (or buffer) and keep the extraction column moist
3-5 ml of non-polar solvent
Leaching extraction column
Rinse the extraction column with 3-5 ml of deionized water/or low ionic strength buffer
Loading
Make the sample at the right flow rate
Uniformly passing through the extraction column
Pass the sample evenly through the extraction column at a lower flow rate
Cleaning
The extraction column of the adsorbed sample is eluted with 5 ml of a solvent of the appropriate polarity (commonly used is pure water, buffer, water/organic solvent mixture)
Extraction column for eluting the sample with 5 ml of the appropriate non-polar solvent
5 ml deionized water / or low ionic strength buffer rinse column
Elution
The sample to be tested is eluted in batches with 1-5 ml of non-polar solvent to the collection container.
The sample to be tested is eluted in batches with 1-5 ml of polar solvent to the collection container.
The sample to be tested is eluted into the collection container with 1-5 ml of high ionic strength buffer.